Usage patterns of the ESCRS IOL Calculator

January 2024 – December 2025

Eligible calculations
2,761,566
Distinct biometric profiles
1,612,637
Website sessions
1,451,859
Countries and territories
203

Loading figures…

Methods

Data sources

Retrospective descriptive study, January 2024 through December 2025. Google Analytics recorded 1,451,859 sessions in its monthly report and 1,455,840 in its separate geographic report. The calculator database recorded one calculation per eye. The two datasets shared no identifier, so calculations could not be assigned to a session, user, or country.

Eligibility and repeat submissions

Of 2,921,411 eye-level entries, 2,761,566 met the requirements for a selected IOL, a supported mode, target refraction, axial length, keratometry, and anterior chamber depth.

Biometry counted identical same-side profiles once across both years. Matching required agreement on available measurements to four decimal places and on which measurements were missing. Age and gender were counted once per submission contributing new biometry. Lens selection and refractive planning retained every eligible simulation.

Measurement ranges
Axial length10–45 mm
Flat and steep keratometry20–110 D each
Anterior chamber depth>0 to 7 mm
Lens thickness>0 to 8 mm
White-to-white distance5–18 mm
Central corneal thickness250–1,000 µm
Target refraction−10 to +10 D
Surgically induced astigmatism0–3 D
Incision location0–360°

Missing or out-of-range optional measurements did not exclude an otherwise eligible entry.

IOL classification

Lens properties were assigned from IOLCon and reviewed by one investigator. Functional categories used the updated ESCRS depth-of-field terminology. Full-DOFi was reported as one group because independent defocus curves were unavailable for some lenses.

Interpretation

Surgeons may have used the calculator selectively for toric or presbyopia-correcting lenses. The records describe choices considered before surgery. Implantation and postoperative outcomes were unknown.

Biometric matching could merge distinct eyes or count an eye again after a measurement changed. The frequency of teaching and test entries was unknown. Geographic rates count repeat sessions and use 2023 workforce estimates.

Authors

Miguel Raimundo, Dante Buonsanti, Adi Abulafia, Giacomo Savini, Damian Clarke, Filomena Ribeiro, and Oliver Findl.

The study received no funding. No ethics committee determination was obtained.